== My MAX31855 output is really unstable and noisy ==
 
== My MAX31855 output is really unstable and noisy ==
   −
If I touch or move the probe the temperature output goes crazy
+
{{bloc-etroit|text=If I touch or move the probe the temperature output goes crazy
The MAX31855 is surprisingly sensitive, we've found a good way to fix this is to place a 0.01uF to 0.1uF capacitor across the thermocouple leads (that is, place the capacitor into the blue terminal block, or solder to the bottom as shown below).
+
The MAX31855 is surprisingly sensitive, we've found a good way to fix this is to place a 0.01uF to 0.1uF capacitor across the thermocouple leads (that is, place the capacitor into the blue terminal block, or solder to the bottom as shown below).}}
    +
[[Fichier:Thermocouple-FAQ-01.jpg|480px]]
    